Agilent Skimmer Cone ICP-MS 7500a, 7500i and 7500s
To fit as standard Agilent ICP-MS 7500a, 7500i and 7500s Sampling and Skimmer Cones The Nickel sampling cone is used as the standard interface. This sampling cone is relatively robust to samples containing high concentrations of matrix and salts, and is suitable for a wide variety of samples. It is usually used with the Nickel skimmer cone. The Platinum sampling cone is usually used for introduction of hydrofluoric acid and other reagents. It is usually used with the Platinum skimmer cone, and is suitable higher concentrations of acids. The Platinum skimmer cone must be mounted on the brass skimmer base(G1820-60104), otherwise the copper base of the skimmer cone will be deteriorated, and shorten the lifetime of the skimmer cone itself.